14 pages Bags versus: a comparison of cell culture systems for the production of dendritic cell–based immunotherapies — Transfusion, 2018

Overview

Review comparing cell culture systems (bags vs flasks) for dendritic cell production. Discusses gas permeability and materials. References Wilson Wolf patents related to gas-permeable devices as examples of improved culture methods utilizing silicone rubber membranes.

What this paper found

This review compares polystyrene flasks with closed fluoropolymer bags for producing monocyte-derived dendritic cell immunotherapies and concludes that vessel choice can affect product quality. Most studies comparing bag and flask cultures found no marked difference in the dendritic cells generated, although some reported differences in CD1a expression or in IL-10 and IL-12 production. Fluoropolymer bags are closed, flexible and gas-permeable, which lowers contamination risk, whereas polystyrene has an elastic modulus of about 3 GPa compared with roughly 500 MPa for fluorinated ethylene propylene. Sipuleucel-T, the first approved cell-based cancer vaccine, was made in fluoropolymer bags and gave a 4-month survival benefit in the phase III IMPACT trial.

Note: This is a literature review; the authors note that systematic studies characterizing DCs from adherent and suspension cultures in different bag containers are still needed.

Selecting the appropriate cell cultivation system is thus fundamental in the design and development of cell production strategies. — Natalie Fekete, Ariane V. Béland, Katie Campbell, Sarah L. Clark, Corinne A. Hoesli, Transfusion

Study details

Authors
Natalie Fekete, Ariane V. Béland, Katie Campbell, Sarah L. Clark, Corinne A. Hoesli
Institution
Department of Chemical Engineering, McGill University, Montreal, Canada · Saint-Gobain Ceramics & Plastics, Inc., Northboro R&D Center, Northborough, Massachusetts
Journal
Transfusion
Year
2018
PMID / PMC
PMID: 29672857
Cell type(s)
DC Cells
Application area
Solid tumor

Explore related

Full citation

Fekete N, Béland AV, Campbell K, Clark SL, Hoesli CA. Bags versus flasks: a comparison of cell culture systems for the production of dendritic cell–based immunotherapies. Transfusion. 2018;58(7):1800-1813.

Building a similar workflow?

Talk to a ScaleReady specialist about how the G-Rex® platform can support your program — from process development to clinical-scale manufacturing.

Start a conversation →