Single-Use, Metabolite Absorbing, Resonant Transducer (SMART) Culture Vessels for Label-Free, Continuous Cell Culture Progression Monitoring

Overview
This article details a single-use metabolite absorbing resonant transducer (SMART) for continuous cell culture monitoring. Polyacrylate films absorb secreted metabolites, altering sensor permittivity. G-Rex®10 static culture vessels are explicitly referenced for validating sensor sensitivity to live cells.
What this paper found
A resonant sensor coupled to a polyacrylate film that absorbs secreted metabolites can track cell growth continuously and non-destructively inside a closed culture vessel. Sensing cells through their intrinsic permittivity gave only a 1.4 MHz shift, whereas the polyacrylate-based SMART dish gave over 30 MHz during four days of HeLa culture, more than a 20-fold sensitivity gain. The film softens as metabolites containing hydroxyl and alkenyl groups, such as terpenoids, accumulate, changing local permittivity. Responses scaled with seeding density (17, 21 and 27 MHz at 30,000, 60,000 and 120,000 cells per cm²), tracked K562 growth, detected niclosamide and marimastat effects, and worked with HEK293, K562, Jurkat, CHO and E. coli.
Note: Authors note that thicker-walled or metal culture vessels may limit electromagnetic coupling between the sensor and the reader antenna.
This indicates that the sensor can dynamically track both progression and cessation of proliferation. — Yee Jher Chan, Dhananjay Dileep, Samuel M. Rothstein, Eric W. Cochran, Nigel F. Reuel, Advanced Science
Study details
- Authors
- Yee Jher Chan, Dhananjay Dileep, Samuel M. Rothstein, Eric W. Cochran, Nigel F. Reuel
- Institution
- Chemical and Biological Engineering, Iowa State University, Ames, Iowa, USA · Skroot Laboratory Inc, Ames, Iowa, USA
- Journal
- Advanced Science
- Year
- 2024
- G-Rex® model(s)
- G-Rex®10
- Cell type(s)
- Other
- Application area
- Other
- Study phase
- Preclinical
How G-Rex® was used
A resonant sensor placed in a G-Rex®10 static breathable flask was used to test K562 cells. In a resuspension experiment with a fully expanded K562 culture, the tall vessel provided enough liquid height to limit the signal from dispersed cells; resuspension dropped the resonant frequency by 1.4 MHz, and it recovered over about 90 min as cells settled. K562 cells were then cultured in the G-Rex®10 for 4 days with the G-Rex® membrane (200 μm) as the sensor insulator, which gave a response only 0.9 MHz above the no-cell control because of background drift from humidity and temperature changes in the incubator.