SMART sensors for cell growth monitoring in closed system G-Rex® for scalable, cost-conscious cell and gene therapy manufacturing

Overview
This article describes integrating a SMART sensor into the G-Rex® bioreactor for real-time, noninvasive monitoring of cell expansion. G-Rex® devices, including the G-Rex® 100M and 500M, are explicitly referenced as the core manufacturing platform for scaling CAR-T cell production.
What this paper found
A single-use resonant sensor built into the G-Rex® culture vessel (SMART G-Rex®), read wirelessly from outside, tracked T-cell expansion without opening the vessel. Four replicate sensors on one vessel showed a maximum coefficient of variation below 1.25%. Across 45 sensors seeded with 50 million activated cells from different donors, a final Skroot Growth Index above 7.08% indicated, with 99% confidence, more than 3.02 billion cells, a greater than 60-fold expansion. Scanning 11 vessels on the benchtop reader took 14 minutes 45 seconds, against 29 minutes 34 seconds for manual sampling. A harvest rule requiring SGI above 7% and a rate of change below 0.05 percentage points per hour flagged only the Day 7 vessel as ready. Three-year accelerated aging did not impair sensor performance.
Note: The harvest thresholds were set for one process; the authors state they need to be tuned for other cell types, media and seeding or activation protocols.
The cell work showed the ability to detect different growth rates, provide an indication for harvest timing, and the potential to eliminate the need of a cell count prior to cryostorage of drug product. — Subhanwit Roy, Samuel Rothstein, Aidan Shervheim, Yee Jher Chan, Cameron Greenwalt, Nathan Neihart, Jose Juarez, Patricia Turner, Andrew Lewis, Nathan Dunkerley, Dan Welch, John Wilson, Dan Fick, Nigel F. Reuel, Cytotherapy
Study details
- Authors
- Subhanwit Roy, Samuel Rothstein, Aidan Shervheim, Yee Jher Chan, Cameron Greenwalt, Nathan Neihart, Jose Juarez, Patricia Turner, Andrew Lewis, Nathan Dunkerley, Dan Welch, John Wilson, Dan Fick, Nigel F. Reuel
- Institution
- Skroot Laboratory Inc., Ames, Iowa, USA · CellReady LLC, Houston, Texas, USA · Wilson Wolf Manufacturing LLC, St. Paul, Minnesota, USA
- Journal
- Cytotherapy
- Year
- 2025
- PMID / PMC
- PMID: 41689922
- G-Rex® model(s)
- G-Rex®100M, G-Rex®500M
- Cell type(s)
- CAR T Cells
- Application area
- Hematologic malignancy
- Study phase
- Preclinical
How G-Rex® was used
Cryopreserved PBMCs were thawed, rested for 4 hours in a G-Rex® 500M brought to 1000 mL, then activated there for 48 to 72 hours with soluble CD3 (0.2 μg/mL) and CD28 (1.0 μg/mL) antibodies to create a bulk of activated cells. The activated cells were seeded into G-Rex® 100M vessels fitted with SMART sensors at 200, 50 or 20 million cells (fast, medium and slow growth) and topped up with 1 L of T cell medium containing 5% HAB serum and 10 ng/mL IL7 and IL15. Cultures expanded for 4 to 8 days; the 50-million-cell condition gave 3.43 to 6.70 billion viable cells.
Explore related
Full citation
Roy S, Rothstein S, Shervheim A, et al. SMART sensors for cell growth monitoring in closed system G-Rex for scalable, cost-conscious cell and gene therapy manufacturing. Cytotherapy. 2026;28(4):102034.
Building a similar workflow?
Talk to a ScaleReady specialist about how the G-Rex® platform can support your program — from process development to clinical-scale manufacturing.